833 Series Data Sheet X-Ray Image Intensifier Power Supply 5 Fully Regulated High Voltage Outputs |
![]() |
| Model 833 Series |
|---|
24V nominal input
Vacuum condition monitor output
Good line and load regulation
Low output ripple and noise
Compact size
Proof against output short circuits and arcing
Remote programming of G1,2 & 3 outputs
High voltage components fully encapsulated| Description |
|---|
|
| Ordering Information & High Voltage Outputs |
||||
|---|---|---|---|---|
| Input Voltage | Output Voltage | Output Current | Control | Order Code |
| 24 V dc | G1: 75V to 350V | 3ľA | 0 to +10V | 833 |
| G2: 400V to 1200V | 1ľA | 0 to +10V | ||
| G3: 2.75kV to 10.3kV | 1ľA | 0 to +10V | ||
| G4: 2.5kV to 3kV | 10ľA | none | ||
| G5: 30kV ( +/-150) | 3ľA | none |
| Electrical Specification | |
|---|---|
| Input | |
| Voltage range | +20.4V to 26.4V dc |
| Line ripple | Up to 0.5V pk/pk |
| Regulation | +/-0.05%, between 20.4V and 26.4V |
| Current (nominal) | 220mA max outputs @ full load |
| Current (Maximum) | 300mA |
| Output (G1,2,3 & 5) |
|
| Voltage range | see Ordering Information and high voltage outputs |
| Current | see Ordering Information and high voltage outputs |
| Control | (G1,2 & 3)Outputs are proportionally controlled, each by an analogue
input signal, 0 to +10V, with an accuracy of +/-2% of demand.
The control input impedance is >10kohm. The CMRR is > 40dB with a CM range +/-0.5V. The HV range is internally limited to
115% of maximum rating if over driven. (G5) None |
| Regulation | +/-0.3 (load) |
| Ripple | <0.3% pk-pk, over line and load range |
| Temperature coefficient | <+/-0.05% per degC between +5 and +55 degC |
| Drift | <+/-0.5% over 8 hour |
| Response time(power on) | 5% in <3 seconds range change: G1/G2/G3, <1 second, min to max demand |
| Output (G4) | |
| Regulation (load) | 2.5kV to 3.0kV output @ <10ľA load |
| Ripple (lf & hf) | <1% pk-pk, @ <10ľA load |
| Protection | Proof against output short circuits of any duration |
| Vacuum condition monitor | |
| Threshold condition | When G4= 1ľA, (+/-20%) |
| Output | Open collector, 30V/20mA max |
| Load (typical) | External LED & resistor to positive rail |
| Status | Open: if I (G4) <1ľA Closed: if I (G4) >1ľA |
| Protection | All HV outputs are protected against intermittent flash-over and continuous short circuit to ground without damage to the HV power supply. The G5 is protected from exceeding 32kV and will shut-down the power supply. An internal, factory fitted (UL) fuse-link protects against failure on the 24V rail or reverse polarity connection. |
| Environmental Specification | |
|---|---|
| Operating temperature range | +5 to +55 deg.C |
| Storage temperature range | -10 to +70 deg.C |
| Humidity | 95%, +55 degC Non condensing |
| MTBF | 35,000 hours |
| Mechanical Specification | |
|---|---|
| Dimensions | 160 x 52.5 x 115mm (6.3" x 2.1" x 4.5") |
| Packaging | Aluminium case, black anodized, with resilient encapsulation. |
| Input connector | 10 way ribbon connector |
| Outputs | 5 off of type BB-0512-50 HV connector (mating half supplied) |
| Mounting | Four M3 X 6mm deep inserts on 140 x 102 mm centres. |