477 Series Data Sheet Laboratory Photomultiplier Supplies Output Voltages: 500V, 1kV ,2kV & 3kV 4 Watts Power Rating |
![]() |
| 477 Series |
|---|
Mains input 115/230V 50/60Hz
Excellent line and load regulation
Low output ripple, noise and temp-co.
Local or remote control
Switch selectable polarity
Voltage and current monitor
Auxiliary split rail supply
Direct reading DPM for voltage & current| Description |
|---|
|
| Ordering Information |
|||||
|---|---|---|---|---|---|
| Input Voltage | Output Voltage | Polarity | Output Current | Ripple pk-pk | Order Code |
| 115/230V ac | 0 to 500V | Switchable | 8mA | 5mV | 477-301 |
| 115/230V ac | 0 to 1kV | Switchable | 4mA | 4mV | 477-302 |
| 115/230V ac | 0 to 2kV | Switchable | 2mA | 2mV | 477-303 |
| 115/230V ac | 0 to 3kV | Switchable | 1.3mA | 5mV | 477-304 |
| Electrical Specification |
|
|---|---|
| Input voltage range | 115/230V ac nominal +/-10%, 50/60Hz (Switchable 2 ranges). |
| VA rating (at full load) | 35VA. |
| HV output voltage & current | See ordering information. |
| HV output ripple | See ordering information. |
| Polarity | See ordering information. |
| Load regulation | <20ppm (0.002%) 0% to 100% load. |
| Line regulation | <20ppm/V (0.002%) over the full input voltage range. |
| Stability (after hr. warm-up) | <15ppm during any 15 minute period and <50ppm during any 8 hour period. |
| Voltage control: | Local: Multi-turn potentiometer allows full control of the output voltage setting. Remote: A signal of 0V to 10.0V corresponds to 0% to 100% of output rating (Input Z >10kohm). Limit dV/dt: Selectable from the front panel ("Ramp"), this limits the rate of change in output voltage to <2V/mS (at max output voltage) reducing the chance of arcing within PM tube. |
| Temperature coefficient | <50ppm (0.005%) / deg.C at max voltage output. |
| Protection | Current is limited to approximately 110% of the maximum rated output, until overload condition is removed. |
| Voltage & Current Monitor | Direct reading 4.5 digit meter, switch selectable between the HV output voltage & current. Also available are buffered monitor outputs for both high voltage output current and voltage (available from the analogue port on the rear panel, 0 to 10V corresponding to 0 to 100% of the rated output). |
| Auxiliary Output | Fixed +/-5 and +/-15V dc @ 250mA, regulation: +/-1%, ripple: 10mV pk-pk. |
| Safety | Designed to meet EN 60950 and UL 1950. |
| Environmental Specification | |
|---|---|
| Operating temperature range | -10 to +50 deg.C |
| Storage temperature range | -10 to +70 deg.C |
| Humidity | 95%, +50 deg.C Non condensing |