149 Series
Data Sheet Digital High Voltage Meter 3 Ranges: 2kV, 20kV & 40/60kV DC Positive and Negative Reading |
![]() |
| Model 149 |
|---|
4.5
digit LCD display
High input impedance
Positive and negative reading without switching
Voltage coefficient compensated
Accuracy better than +/-0.05%
Precision output for recorder or oscilloscope
40kV high voltage probe supplied with 149-01
Rugged construction| Description |
|---|
There are two meters in the range, the first meter offers voltage measurement up to 40kV and the second meter offering measurement up to 60kV. Both offer high input impedance drawing <10ľA at their maximum input voltages and have three selectable ranges covering : 0-2kV with 0.1V resolution 0-20kV with 1V resolution and 0-40kV / 60kV with 10V resolution. |
| CAUTION |
|---|
Great care must be taken when making measurements in high voltage circuits. ALWAYS ensure that the HV Return terminal of the meter is grounded. DO NOT use hand-held probes on supplies that are potentially lethal. |
| Ordering Information |
|||
|---|---|---|---|
| Range(s) | Accuracy | Input resistance | Order Code |
| 2kV, 20kV & 40kV | +/-0.1% | 10 Gohm | 149-01 |
| 2kV, 20kV & 60kV | +/-0.05% | 10 Gohm | 149-02 |
| 2kV, 20kV & 40kV | +/-0.1% | 100 Gohm | 149-03 |
| 2kV, 20kV & 60kV | +/-0.05% | 100 Gohm | 149-04 |
| Electrical Specification |
149-01 & 149-03 | 149-02 & 149-04 |
|---|---|---|
| Voltage ranges | 0 to 2kV dc, 20kV dc & 40kV dc. | 0 to 2kV dc, 20kV dc & 60kV dc. |
| Resolution | 0.1V on 2kV range, 1V on 20kV range | 0.1V on 2kV range, 1V on 20kV range and 10V on 40kV range. and 10V on 60kV range. |
| Accuracy | +/-0.1% on the 2kV, 20kV & 40kV ranges | +/-0.05% on the 2kV, 20kV & 60kV ranges |
| Input resistance |
149-01 & 149-02
10 Gohm on all ranges | |
| Stability | Better than 0.3% over 1 year. | |
| Polarity | Positive or negative, auto reading. | |
| Protection | Circuits protected against damage by external arcing, flashover etc. | |
| Recorder output | 10mV per
kV (Source/sink capability up to 100ľA.) Accuracy +/-0.25% at 30kV (+/-0.5% at the extremes). Initial offset <0.2mV. Thermal drift <150 ppm per deg.C. Long term drift <0.5% over 1 year. | |
| Controls | Push button power On / Off and range selection. | |
| Battery type | Internal battery, compartment accessible from the rear panel (compartment containing 2 off type "PP3" 9V batteries, alternatively Ni Cd or equivalent cells may be used). | |
| Battery life | Approx. 100 hours of continuous operation from an alkaline battery. | |
| Battery check | Automatic: "BAT" shown on LCD display when battery is low. | |
| Environmental Specification | |
|---|---|
| Operating temperature range | 0 to +50 deg.C |
| Storage temperature range | -20 to +65 deg.C |
| All high voltage components are fully encapsulated for protection against humidity. | |